• myosist
  • myosist

Myosist

Intra Aortic Balloon Catheter Kit (8F)

The MYOSISTTM Intra Aortic Balloon Catheter Kit & accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.

Device Placement : 

  • The device is placed in the descending aorta approximately 2 centimeters (0.79 in) from the left subclavian artery.
  • By balloon inflation in aorta during diastole and deflation during systole, it increases blood supply to heart muscle and decreases work of  the left ventricle.

Benefits

myosist
  • 01

    Highly Durable & Flexible Balloon Membrane 
    Tested for 1 million inflation-deflation cycles for functionality and performance integrity.

  • 02

    Radiopaque Catheter Tip & Markers 
    Easy to locate the MYOSISTTM deployment under fluroscopy. Catheter consists of optimized placement of radiopaque markers and radiopaque tip

  • 03

    Optimized Insertion Length 
    With 710 mm insertion length, the MYOSISTTM is placed in proper positioning.

  • 04

    Inner Lumen Diameter 
    0.028 inch diameter provides a medium to measure excellent arterial pressure

Indication

  1. Cardiogenic shock :

    • Associated with acute MI, mechanical complications of MI - MR , VSD

  2. In association with CABG :

    • Preoperative insertion
    • Patients with severe LV dysfunction
    • Patients with intractable ischemic arrhythmias
    • Postoperative insertion
    • Postcardiotomy cardiogenic shock
  3. In association with nonsurgical revascularization:

    • Hemodynamically unstable infarct patients
    • High risk coronary interventions
    • severe LV dysfunction, LMCA, complex coronary artery disease
  4. CVTS In Cardiopulmonary Bypass:

    • Coronary bypass Graft
    • Cardiac valve repair (Aortic, mitral, tricuspid, pulmonary)
    • Repair of large septal defects (atrial septal defects, venticular septal defects)
    • Transplantation (Heart and Lung). 

 

Product Specifications

Balloon Volume (cc) 30 35 40
Guideline Patient Height (cm) <152   152-162 162-183
Balloon Diameter (mm) 14.5 15 15
Balloon Length (mm) 225 227 259
Balloon Catheter Length (mm)   855  
Balloon Catheter Usable Length (mm)   710  
Catheter Profile (french)   8  
Inner Lumen ID (inch)   0.028  
Guidewire Size (inch)   0.025  
Console Compatibility Datascope®: System 97, 98, 98XT, CS100, CS300, CARDIOSAVE
Arrow®: ACAT, AutoCAT, AutoCAT 2, AutoCAT 2 Wave, KAAT II

Size Chart

Product Code Product Code Description
MYS0830 Myosist IAB Catheter Kit 8F 30CC
MYS0835 Myosist IAB Catheter Kit 8F 35CC
MYS0840 Myosist IAB Catheter Kit 8F 40CC