- Next-generation automated biochemistry system to meet the needs of high-end laboratories.
- Closed system reagents with calibrators and controls with viable software.
- Capability of handling a workload of a medium to high, throughput laboratory.
- Includes 400 tests/hour throughputs.
- Potential to provide cost-effective solutions for day-to-day as well as specific testing parameters.
Key Features
FAQs
Product Details
Technical
Specifications
Item | Description |
Sampling Unit | |
System Type | Turn Table type tray |
Smart positions | 75 positions |
STAT samples | Any position |
Sampling probe | Sample volume: 2.7–5 μL (0.1 μL steps); Sample replication possible for same test |
Sampling drive | Micro-pipette with Y-axis sensor; External & internal washing with de-ionized water |
Sample Identification | Equipped with Hematocrit & hemolysis detection facility |
Sample range | Dilution: x2 to x40 for auto dilution |
Sample ID | Barcode for sample details and position |
Reagent Unit | |
Tray type | Turn table type R & R |
Reagent positions | 60 positions |
Cooling | 6–15°C (with built-in real-time refrigeration unit) |
Reagent volume | 180–500 μL (depending on reagent size) |
Reagent container | 20 mL (sealed bottle with identification system, external & internal washing with de-ionized water) |
Reagent type | Reagent 1: 180–570 μL adjustable (in 1 μL steps); Reagent 2: 0 or 10–200 μL (adjustable in 1 μL step) |
Reagent identification | Barcode or ID based |
Water tank condition | Wash with System Water / Select Reagent / Wash Solution |
Reaction Unit | |
Reaction type | Turn Table |
Cuvettes | 80 Reaction Cuvettes |
Reaction temperature | 37.0 ± 0.2°C |
Incubation | 30–600 seconds |
Optical System | |
Type | Multiple wavelength; Spectrophotometer with fiber optics cable & sealed filters |
Light source | Tungsten Halogen lamp |
Filters | 8 Optical Filters (340, 405, 450, 505, 546, 578, 620, 670, 700 nm) |
Calibration & Quality Control | |
Calibration | Max 20 calibrators; 6 calibrators maximum per chemistry |
Calibration curve | Choice from linear regression, Auto select, Spline, Logit, or concentrated calibrator |
Quality control | Westgard rules (up to 6), L-J chart, twin plot, X-R control diagram, MA (Graphs) |
Process Management | |
Reflex | Reflex order can be programmed by defining panic ranges |
Offline (PC mode) | Offline: Planning orders, Calibration curve and patient data |
Online | Online: Real-time display & result validation |
Data storage capacity | 80,000 Patient tests; 40,000 QC tests |
Report (short/long) | Patient, Test wise, Test date, etc. |
Alarms | Sample/reagent low level; Waste container full; Warnings on display |
Printer | Graphic thermal built-in (optional external) |
Working Condition | |
Power supply | 110/220 VAC, 50/60 Hz (Max. Power: 250 W) |
Room temperature | 15°–30°C; Humidity: ≤ 85% |
Water | De-ionized water with temperature less than 2°C per hour |
Drain | 30–80 cm from water formation |
Dimensions | 78 (W) × 66 (D) × 60 (H) cm |
Approx. weight | 75 kg |